Infinium® II Assay HC Two-Sample, AutomatedExperienced User CardPart # 11280811 Rev. A Page 1 of 22Day 1 Day 2 Day 3Pre-AmpPost-AmpOptional stop,cold
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 10 of 22[ ] 18. Place the inverted, uncovere
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 11 of 22Start the Robot[ ] 1. At the robot P
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 12 of 22Estimated TimeRobot time:• 8 samples:
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 13 of 22[ ] 1. Place the Hyb Chamber gaskets
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 14 of 22Set Up HC BC2 for Hyb[ ] 1. Ensure t
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 15 of 22Estimated TimeRobot time:• 4 BeadChips
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 16 of 22StepsWash BC2[ ] 1. Attach the wire
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 17 of 22[ ] a. Gently push the glass back pl
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 18 of 22Estimated TimeRobot time: • 2 hours an
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 19 of 22[ ] Turn on the water circulator. Se
Infinium® II Assay HC Two-Sample, Automated (Pre-Amp)Experienced User CardPart # 11280811 Rev. A Page 2 of 22Estimated TimeRobot time:• 20 minutes pe
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 20 of 22[ ] 2. When prompted, wait for the C
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 21 of 22[ ] 7. Move the BeadChip staining ra
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 22 of 22[ ] 4. In the Settings area, click E
Infinium® II Assay HC Two-Sample, Automated (Pre-Amp)Experienced User CardPart # 11280811 Rev. A Page 3 of 22[ ] 4. Vortex each WG#-DNA plate:• For
Infinium® II Assay HC Two-Sample, Automated (Pre-Amp)Experienced User CardPart # 11280811 Rev. A Page 4 of 22Estimated TimeRobot time: 5 minutes per
Infinium® II Assay HC Two-Sample, Automated (Pre-Amp)Experienced User CardPart # 11280811 Rev. A Page 5 of 22Estimated TimeRobot time:• 20 minutes pe
Infinium® II Assay HC Two-Sample, Automated (Pre-Amp)Experienced User CardPart # 11280811 Rev. A Page 6 of 22[ ] 4. (Non-Infinium LIMS only) Make s
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 7 of 22Estimated TimeRobot time:• 10 minutes f
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 8 of 22[ ] 2. Click OK in the message box. R
Infinium® II Assay HC Two-Sample, Automated (Post-Amp)Experienced User CardPart # 11280811 Rev. A Page 9 of 22[ ] 3. Centrifuge the sealed AMP2 pla